TM-ABI Technical Specifications

Power & Battery

Compliant with standards:
EN 60601-1:2007 General requirements for basic safety and essential performance

EN 60601-1-2:2008 Electromagnetic compatibility requirements and tests
EN 80601-2-30:2010 Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers

Flammable anaesthetics: WARNING! Not suitable to use with flammable anaesthetics.

AC/DC power supply: FRIWO FW7556M
Input: 100-240V AC/50-60Hz/350mA
Output: 5V DC/3.0A

Battery type: rechargeable lithium polymer
Capacity: 2,300mAh
Number of measurements per charge: 50

Electrical specifications:
100-240V Ac, 50-60Hz, 3A
Type of protection against electric shock:
class II

Other specifications

Cuff inflation and deflation
Automatic inflation using an air pump and deflation using an electromagnetic valve.

Temperature and humidity range
Working environment: 10 to 40°C, 30 to 85% relative air humidity
IPX0 protection, transport and storage: 0 to 60°C up to 85% relative air humidity

Measure specifications

Measurement types:
Ankle-brachial index using the oscillometric method
Systolic blood pressure using the oscillometric method
Diastolic blood pressure using the oscillometric method
Heart rate using the oscillometric method
Oscillometric method is one of plethysmographic techniques.

Measurement range:
Pressure: 0 to 299 mmHg
Heart rate: 30 to 199 BPM
Limit values of measurement errors:
Blood pressure: Within ± 3 mmHg or 2%
Pulse rate: Within ± 5% of BPM
ABI: Within ± 0.1

Reliable even in the case of critical ischemia and medial calcinosis

It is crucial to provide a reliable measurement also when examining a patient with severe PAD.
Our improved plethysmographic sensors detect critical ischemia and medial calcinosis even when pressure oscillations are not available due to heavy occlusion.
Every measurement with TM-ABI provides sufficient information for further actions.